| ULTRA HIGH-SPEED TRANSIENT DIGITIZERS

SCD SERIES TRANSIENT WAVEFORM RECORDERS
- High single-shot sample rate to 200 GS/s
- High analog bandwidth
- 4.5 GHz with SCD5000
- 1 GHz with SCD 1000
- Input sensitivity flexibility with SCD1000
- Time resolution to 5 ps per point
- Variable record length from 256 to 1024 points
- Display unit for easy control, viewing, and measurements
APPLICATIONS
- Laser measurements
- Accelerator diagnostics
- EMP and radiation testing
- Nuclear effects
- Telecommunications
- EMC/EMI measurements
- Radar
HIGH PERFORMANCE CAPTURE
The SCD5000 and SCD1000 are designed to capture low nanosecond and
picosecond single-shot events. Whether the application involves Laser, ESD, EMP, Accelerators or other high-speed single-shot phenomena, the
SCD series waveform recorders can capture the event with excellent
fidelity. The maximum acquisition rate of 200 GS/s provides time resolution to 5
ps. With time windows from 5 ns to 100 µs, the SCD series recorders
provide flexible acquisition windows. When the delay line is used,
approximately 2.5 ns of pretrigger information can be captured.
The SCD5000 uses direct access with input sensitivity of 5 V and offset
of ±4.0 V. The SCD1000 provides quality conditioning of the input signal
with 100 mV to 10 V full scale ranges, dc offset and ac or dc coupling.
There are two input channels which can be configured to multiplex two
signals or algebraically add or subtract two input signals to the single
channel waveform recorder.
The waveform record length is selectable between 256, 512 or 1024 data
points. With 16 built-in waveform locations, multiple trigger events can
be stored into separate storage locations. Each record is time and date
stamped for later comparison. There are 10 nonvolatile settings storage
locations for quick instrument setup.
STAND-ALONE OR SYSTEM SUPPORT
The SCD Series recorders are designed for use in either a stand-alone or
a system environment. With the standard display unit, SCD waveform
recorders allow full control of operation, viewing of up to 4 waveforms
and cursor measurements.
In a system environment, there is a high-speed GPIB interface and
commands to make data acquisition flexible and fast.
OPTIONS EXPAND CAPABILITY
A fast waveform capture option (Opt, 1 P) increases the capture rate by
ten times, from one waveform per second to ten waveforms per second.
There is a choice of input connectors in addition to the the standard
Type-N connector. On the SCD1000, a choice of SMA or Tek Type II
intelligent probe interface are available. On the SCD5000, SMA
connectors can be chosen.
CHARACTERISTICS
The characteristics are divided into 3 parts:
• Vertical characteristics for all SCD series recorders
• Trigger characteristics for all SCD series recorders
• SCD series common characteristics
| VERTICAL |
|
|
|
| Parameter |
SCD1000 |
SCD5000 |
SCD5000 Opt 01 |
| Bandwidth |
dc coupled; dc to 1.0 GHz;
ac coupled; 100 kHz to 1.0 GHz. |
4.5 GHz |
3.0 GHz |
| Rise time |
350 ps |
80 ps |
120 ps |
| Input Range |
100 mV to 10 V full scale input in a 1, 2, 5 sequence (7 steps) |
±2.5 V direct access |
±5Vdirect access |
| Offset |
±2.5 times input range (peak to peak) |
±4 volts |
±4 volts |
| Invert |
Invert on A & B channels |
No |
No |
| Input Impedance |
DC coupled; 50Ω, AC coupled; 50Ω in series with 2.2
μF |
50Ω, DC coupled |
50Ω, DC coupled |
| Coupling |
AC and DC |
DC |
DC |
| Maximum Input Voltage |
5 Vrms (0.5 W) or 0.25 W-sec. Pulses not exceeding 25 V peak |
5 Vrms (0.5 W) or 0.25 W-sec. Pulses not exceeding 25 V peak |
5 Vrms (0.5 W) or 0.25 W-sec. Pulses not exceeding 25 V peak |
| VSWR |
|
Less than 1.5, dc to 3.5 GHz |
|
| Vertical resolution |
11 bits |
11 bits |
11 bits |
| TRIGGERING |
|
|
|
| Parameter |
SCD1000 |
SCD5000 |
SCD5000 Opt 01 |
| Source |
CH A, CH B, External analog |
External analog |
External and internal analog |
| Coupling |
AC or DC |
AC only |
AC only |
| Slope |
Positive or Negative |
Positive or Negative |
Positive or Negative |
| Level range |
Internal; ±125% of input range in 1 % steps. External analog; ±1.25 V in 10 mV steps. |
External analog; ±1.25 V in 10 mV steps. |
External analog; ±1.25 V in 10 mV steps. |
| Sensitivity (sinewave) External input |
50 mV, 20 kHz to 50 MHz
150 mV, 50 MHz to 500 MHz
350 mV, 500 MHz to 1 GHz |
50 mV, 20 kHz to 50 MHz
150 mV, 50 MHz to 500 MHz
350 mV, 500 MHz to 1 GHz |
50 mV, 20 kHz to 50 MHz
150 mV, 50 MHz to 500 MHz
350 mV, 500 MHz to 1 GHz |
| Sensitivity (sinewave) Internal input |
0.05 x range, DC to 250 MHz
0.15 x range, 250 MHz to 1 GHz |
N/A |
500 mV, 20 kHz to 50 MHz
1.5 V, 50 MHz to 500 MHz
3.5 V, 500 MHz to 1 GHz |
| Sensitivity (pulse) |
150 mV with minimum of 500 ps HAD
(half amplitude duration) pulse. |
150 mV with minimum of 500 ps HAD (half amplitude duration) pulse. |
150 mV with minimum of 500 ps HAD (halt amplitude duration) pulse. |
| Pretrigger view time |
2.5 ns |
N/A |
2.5 ns |
SCD SERIES COMMON CHARACTERISTICS
Time per Point - 5 ps/point (5 ns time window with 1024 point record
length) to 400 ns/point (100µs time window with 256 point record
length).
Time Windows - 5 ns to 100 Ns in a 1, 2, 5 sequence of 14 steps.
Trigger delay - 0% to 500% of time window with 0.06% resolution.
DIGITIZING
Record Length -256, 512 or 1024 waveform data points.
Internal Calibration - Internal signal standards for increased accuracy
of vertical range, dc offset, trigger level, and window size.
Multiple Records - 16 waveform locations (first 4 locations are battery
backed up).
Auto Advance Mode -Automatic capture of sequential events into separate
waveform locations.
Time Stamping - Time and date stored with every
acquisition.
SYSTEM CAPABILITIES
GPIB - IEEE-488.1 interface is standard for instrument control and
waveform data transfer. Maximum transfer rate ≥ 500 kilobytes/s.
Instrument functions, settings, and operating modes are programmable,
with the exception of the power switch.
Waveform Data Formats - Either processed (using centroid algorithm) or raw
waveform data available.
Settings Storage - 10 nonvolatile instrument
settings.
System commands - Examples of commands include:
• Advance data into multiple waveform locations (16)
• NRepeat makes user specified single-shot acquisitions
and automatically transfer to controller
• Calibrate command for increased time and amplitude
accuracy
• Data Count allows partial waveform transfers
DISPLAY UNIT
Compatibility - Attaches to any SCD Series waveform recorder.
Menu Functions - Instrument settings and operating modes are
programmable from the display unit.
Waveform Display - Up to 4 waveforms
displayed.
Local Cursor Measurements - Volts, ΔVolts, Time, ΔTime, and 1/ΔTime
on one waveform or between any two waveforms.
SIGNAL CONNECTIONS
Gate out - Single BNC connector on back panel, output TTL level pulse is
coincident with start of sweep.
Video out -TTL, VGA (600 x 400
resolution) compatible video signal output on back panel. Compatible
with standard PC multiple-frequency monitors and video hardcopy units.
ENVIRONMENTAL (WITHOUT DISPLAY UNIT)
Temperature Range - Meets Mil-T-28800D type III, Class 5. Operating
-0 to +50°C. Non-Operating -55 to +75°C.
Humidity - Meets Mil-T-28800D type III, Class 5. Up to 95% relative
humidity (noncondensing).
Altitude - Exceeds Mil-T-28800D type III, Class 5. Operating: 15,000 ft.
(4.5 km) maximum. Nonoperating: 50,000 ft. (15 km) maximum.
Electromagnetic Compatibility - FCC Part 15, Subpart J, class A and VDE
0871/6.78 class B.
POWER
Line Frequency - Operational from 48 to 440 Hz.
Power Consumption -
SCD5000: ≤250 Watts maximum fully optioned. SCD1000: ≤300 Watts maximum
fully optioned.
Line Voltage Range -90 to 132 Vrms (115 V); 180 to 250
Vrms (230 V).
ORDERING INFORMATION
SCD5000 4.5 GHz Waveform Recorder MSRP: $46,900, MACHTIK PRICE:
$37,500
$27,500 SPECIAL!!
Includes: Standard configuration with detachable display unit and Type N
connectors, power cord, Instruction Manual (070-6960-00), Instrument
Interfacing Guide (070-7315-00) Quick Reference Card (070-7316-00),
rackmount hardware.
SCD1000 1.0 GHz Waveform Recorder MSRP:
$29,900, MACHTIK PRICE: $24,500
Includes: Standard configuration with detachable display unit and Type N
connectors, power cord, Instruction Manual (070-6960-00), Instrument
Interfacing Guide (070-7315-00), Quick Reference Card (070-7316-00),
rackmount hardware
OPTIONS
Opt. 1P - Increase waveform capture rate from one to ten
512-point waveforms per second. MSRP: +$3,000,
MACHTIK PRICE: +$2,500
Opt. 1E -Tek Type II Intelligent Probe interface (SCD1000 only).
MSRP: +$900, MACHTIK PRICE: +$700
Opt. 2E - SMA input connectors. SCD5000 +$200, SCD1000
MSRP: +$250, MACHTIK PRICE: +$200
Opt 01 - Add delay line with Internal trigger pickoff (SCD5000
only). MSRP: +$3,000, MACHTIK PRICE: +$2,500
Opt. 20 - Delete detachable display unit. MSRP: -$2,900, MACHTIK PRICE: -$1,000
SCD FAMILY FIELD UPGRADE
SCDF019 - Add detachable display unit for SCD Series waveform
recorder. MSRP: $3,200, MACHTIK PRICE: +$2,500
SCDF01P - Add fast waveform capture option. MSRP: $3,500, MACHTIK PRICE: +$3,000
SCOF001 - Add delay line to SCD5000. MSRP: $4,000, MACHTIK PRICE: +$3,200
OPTIONAL ACCESSORIES
SCD1000/SCD5000 Service manual Order 070-6963-00 MSRP: $150, MACHTIK PRICE: +$150
GPIB Cable, Double Shielded, Low EMI-1 meter 012-0991-01 MSRP: $140, MACHTIK PRICE: +$100
Type N male to SMA female adaptor - MACHTIK
PRICE: +$50
Type N male to BNC female adaptor - Order 103-0045-00 MSRP: $17, MACHTIK PRICE: +$5
Type N male to GR adaptor - Order 017-0021-00 MSRP: $90, MACHTIK PRICE: +$75
PHYSICAL CHARACTERISTICS
Width: 19 in. - 482.6 mm
Height: 7 in. - 177.8 mm
Depth: 30 in. - 762 mm
Net Weight: 54 lbs. - 24.5 kg |